Yokogawa Hydrofluoric Acid-Resistant pH Sensor HF405
Yokogawa's pH and ORP analyzers are highly reliable. They feature advanced functions and are useful for a wide variety of applications including water quality management in a broad range of production processes and in medium-sized wastewater treatment plants, or as general pH and ORP control systems.
Based on Yokogawa's track record and years of experience, a comprehensive range of products has been produced to provide solutions best suited to individual applications.
In addition to the PH8EFP, PH8ERP, OR8EFG and OR8ERG standard Ryton pH and ORP sensors, specialty pH and ORP sensors are available for various applications.
Features
Yokogawa Hydrofluoric Acid-Resistant pH Sensor HF405
pH/ORP Sensors for Chemical Processes : DPA405, DPA485
• Extremely long life
span for pH measurement in electrolytic processes
• With the pressurized inner solution there is no
need for a pressure holder.
• The silver barrier
incorporated in the reference electrode inhibits the generation of sulfide
around the liquid junction.
Specifications
Yokogawa Hydrofluoric Acid-Resistant pH Sensor HF405
pH Sensors for Chemical Processes (DPA405) Measuring range: pH 0 to 14
Measuring temperature:
0 to 100 OC (See Table 1 when using holder)
Measuring pressure:
Atmospheric pressure to 250kPa (Depending on the
inner pressure of the sensor)
Internal
electrolyte: High-viscosity gel
Temperature compensation sensor:
None (Manual temperature compensation on the
converter or transmitter)
(Use adapter with temperature sensor (SA405) for
application where the tempera- ture varies)
Applicable holder:
Yokogawa Hydrofluoric Acid-Resistant pH Sensor HF405
Flow-through holder (PH8HF), Submersion holder
(PH8HS)
(An optional adapter is needed, but not needed
when using an adapter with temperature sensor SA405.)
Ultrasonic cleaning is not available.
Only jet cleaning is available when cleaning is
necessary.
Use O-ring covered by Teflon (K9148MR) when
using a special holder for electrolytic processes
Wetted part
materials: Body: Glass
O-ring: Silicon rubber or Daielperfrow rubber
Adapter: Stainless steel (SUS316), polypropylene,
rigid polyvinyl chloride, heat-resistant polyvinyl chloride or titanium
CAUTION ON USE:
This sensor cannot be used outdoors, even when
using with a holder, this sensor cannot be used outdoors due to exposure to
rain or due to condensation at a high humid place. This sensor cannot be used
with a guide- pipe holder. The sensor must be installed in a vertical
position. It can not be installed from below and
in a horizontal position, either.