 Detailed description

Yokogawa Terminal Box BA20/YT


GENERAL

The ISC40GJ, ISC40SJ sensors are designed for use with the FLXA21 2-wire Analyzer, ISC202 2-wire transmitter and the ISC450G 4-wire converter. This combination exceeds all expectations for conductivity measurement in terms of reliability, accuracy, rangeability and price performance.

This innovative inductive conductivity sensor provides highly accurate measurements over a wide measuring range (1 μS/cm to 1999 mS/cm) and process temperature range (-10 to 130oC, -10 to 105oC for ISC40SJ-TT) without changing the cell constant and conducting recalibration.

The erosion/abrasion resistant PEEK (Poly Ether Ether Ketone), which also features excellent chemical resistance in all solutions except Fluoric Acid or Oxidizing Concen- trated Acids.

The PEEK sensor is provided with a rugged Stainless Steel mounting thread/nut/gasket combination for ultimate flexibility in installation using bulk head installation technique. There is also a wide range of holders and options available for reliable in-line or off-line installation with double O-ring seals for long service life of the sensor.

The ISC40GJ, ISC40SJ have a large bore for optimal resistance to fouling processes and when properly installed, the flow will keep the sensor clean, to help avoid measuring errors.

 

FEATURES

Yokogawa Terminal Box BA20/YT

• Inductive Conductivity technique for elimination of fouling and polarization errors.

• Wide bore sensors for long term stability.

• Installation flexibility due to wide range of holders and due to the use of universal bulkhead construction.

• A single sensor can maintain the high resolution and accuracy, and measure the conductivity in an extremely brood range.

  Minimum span: 100 μS/cm

  Maximum span: 1999 mS/cm

 

APPLICATIONS

Yokogawa Terminal Box BA20/YT

• All applications where severe electrode fouling prevents the use of contacting electrodes.

• All ranges except (ultra) pure water applications.

• All slurry applications where conventional systems suffer from plugging or erosion.

• All applications where the 6 decade rangeability is necessary for accurate process control.

 

GENERAL SPECIFICATION 

Yokogawa Terminal Box BA20/YT

1. ISC40GJ, ISC40SJ Inductive Conductivity Sensor

   Compatibility:

   ISC40GJ is compatible with FLXA21 2-wire Analyzer, ISC202G Inductive Conductivity Transmitter and ISC450G Converter.

   ISC40SJ-TT is compatible with ISC202SJ Inductive Conductivity Transmitter.

   Measuring range: 1 μS/cm to 1999 ms/cm

   Output span: Minimum 100 μS/cm

                           Maximum 1999 mS/cm

  Process temperature:

      -10 to 130 C for continuous exposure.

      -10 to 105 C for ISC40SJ-TT. Suitable for steam-sterilisable applications.

 Process pressure:

   Dependent on installation; but <2 MPa (300 psi).

   (Note) Process temperature and pressure depend on specification of holders and adapters.

 Process flow: Maximum 5 m/s.

 Wetted materials:

   Sensor: PEEK (Poly Ether Ether Ketone).

   O-ring: Viton® or ethylene propylene copolymer rubber.

   Adapter (optional): SUS316 Stainless steel or PVC or PVDF.

 Non-wetted materials:

    Sensor thread: SUS304 Stainless steel.

    Retaining nut:  SUS304 Stainless steel.

    Cable: Weatherproof vinyl.

 Process connection:

    With retaining nut on G3/4 thread of sensor top (refer to section Drawings and Dimensions)      for bulkhead mounting; optional process adapters      or process fittings.

  Process adapters:

    JIS 10K 50 RF flange adapter (Material: SUS316)

    JIS 10K 50 FF flange adapter (Material: PVC or PVDF)

    DIN PN16 DN50 flange adapter

    ANSI Class 150 2 flange adapter

    R2 screw-in adapter

 Cable length: 5 m, 10 m, 15 m, 20 m

  The length into extension cable is inside of 50 m

  Extension cable cannot be used with ISC40SJ-TT.

 Dimensions: Refer to section Drawings and Dimensions.

 Weight:   Sensor: approximately 0.6 kg.

 (Note)  Do not submerge the sensor itself in process water, as the seams between the mold and the metal of the sensor are not waterproof. Since a temperature sensor is imbedded in the PEEK molded sensor, its response speed is not fast. Install another temperature sensor if necessary.

 ISC40SJ-TT Intrinsically safe type sensor

 TIIS certification sensor should be used with ISC202SJ Protection Concept and Adapter Group: Ex ia IIC T4 Intrinsic safe rating:

Ui=14.4 V, Ii=20 mA, Pi=190 mW, Li=28.6 mH, Ci=0.2 μF

 Environment and operational conditions

 The temperature of the sample solution in contact with sensor should be the range of -10 to 105 C.

 

2. ISC40FDJ Immersion Holder

  Process temperature: Maximum 80 C.

  Process pressure: Maximum 0.2 MPa at 20 C.

                                   Maximum 0.1 MPa at 80 C. Wetted materials:

  Holder: C-PVC or SUS316

  O-ring: Viton® or ethylene propylene copolymer rubber. 

  Flange (Optional) : PP or SUS316

  Gasket: Chloroprene or ethylene propylene copolymer rubber.

  Process connection:

   Fixing flange (Optional): DIN PN10 DN50 (ANSI 2 inch 150 lbs. with bolt holes): Material PP

  JIS 10K 50 RF: Material SUS316 2-inch pipe Mounting set (Optional): Zinc-plated steel.

 

3. ISC40FFJ Flow Holder

  Process temperature:

  ISC40FFJ-SA, ISC40FFJ-SJ:  Maximum 150 C.

  ISC40FFJ-PA, ISC40FFJ-PJ:  Maximum 100 C.

  ISC40FFJ-FA, ISC40FFJ-FJ:  Maximum 130 C.

 Process pressure:

  ISC40FFJ-SA, ISC40FFJ-SJ: Maximum 1.0 MPa at 150 C.

  ISC40FFJ-PA, ISC40FFJ-PJ: Maximum 0.6 MPa at 20 C.

       ISC40FFJ-FA, ISC40FFJ-FJ: Maximum 0.1 MPa at 100 C.

                                                           Maximum 1.0 MPa at 20 C.

                                                           Maximum 0.1 MPa at 130 C.

 Wetted materials:

    ISC40FFJ-S口:    SUS316

    ISC40FFJ-P口:    Polypropylene

    ISC40FFJ-F口:    PVDF

  O-ring:  Viton® or ethylene propylene copolymer rubber.

 Non-wetted materials:

  Nut: SUS304

  Mounting set (Optional): SUS304

  Flange adapters (Optional): SUS304

Process connection:

1/2NPT or Rc1/2

DIN PN10 DN25 flange adapters (Optional)

JIS 10K 25 RF flange adapters (Optional)


4. ISC40FSJ Direct Insertion Subassembly

Process temperature:

  ISC40FSJ-STWJ : Maximum 110 C.

  ISC40FSJ-SCWJ, -SCSJ: Maximum 150 C.

  ISC40FSJ-PCSJ : Maximum 100 C.

Process pressure:

ISC40FSJ-SCWJ, -SCSJ: Maximum 1.0 MPa at 150 C.

ISC40FSJ-PCSJ: Maximum 0.6 MPa at 20 C.

                                 Maximum 0.1 MPa at 100 C.

ISC40FSJ-FCSJ: Maximum 1.0 MPa at 20 C.

Maximum 0.1 MPa at 130 C.

Materials:

Wetted materials:

ISC40FSJ-STWJ: SUS316L, silicon rubber.

ISC40FSJ-SCWJ, -SCSJ: SUS316, Viton® or ethylene propylene copolymer rubber.

ISC40FSJ-PCSJ: Polypropylene, Viton® or ethylene propylene copolymer rubber.

ISC40FSJ-FCSJ: PVDF, Viton® or ethylene propylene copolymer rubber.

Non wetted materials:

       ISC40FSJ-STWJ : IDF clamp: SUS13. 

       ISC40FSJ-SCWJ, -SCSJ, -PCSJ, -FCSJ: Nut: SUS304.

Process connection:

  ISC40FSJ-SCWJ: coupling.

  ISC40FSJ-SCSJ-PCSJ-FCSJ: R2 screw-in coupling.

Dimensions:

  Refer to section Drawings and Dimensions.

 

5. BA20 Terminal Box

Use when ISC202/FLXA21 transmitter or ISC450G converter is separated from ISC40口J sensor and is set up.

Ambient temperature: -10 To 50 C

Constructions: IP54 agreement

Case material: Article of cast metal of aluminum alloy

Cable inlet: 2 (Pg13.5) 

Case color : Straight gray 

Weight : Approx. 2 kg

(Note) BA20 can not be used with ISC40SJ-TT.

 

6. WF10J Extension Cable

Number of mind Line: 6

Finish outside diameter: 7.7 mm

Terminal processing: Special terminals

Material: Weatherproof vinyl.

    (Note) WF10J cannot be used with ISC40SJ-TT.