Product Information

Yokogawa General Purpose Inductive Conductivity Sensor ISC40GJ-GG-T1-10/SSG[Product print page]

Product Name: Yokogawa General Purpose Inductive Conductivity Sensor ISC40GJ-GG-T1-10/SSG
Product Type: ISC40GJ-GG-T1-10/SSG
Exhibitors: Yokogawa
Product Documentation: No Related Documents

Brief

Yokogawa General Purpose Inductive Conductivity Sensor ISC40GJ-GG-T1-10/SSG, Yokogawa General Purpose Inductive Conductivity Sensor ISC40GJ-GG-T1-10/SSG


 Detailed description

Yokogawa General Purpose Inductive Conductivity Sensor ISC40GJ-GG-T1-10/SSG



GENERAL

The ISC40GJ, ISC40SJ sensors are designed for use with the FLXA21 2-wire Analyzer, ISC202 2-wire transmitter and the ISC450G 4-wire converter. This combination exceeds all expectations for conductivity measurement in terms of reliability, accuracy, rangeability and price performance.

This innovative inductive conductivity sensor provides highly accurate measurements over a wide measuring range (1 μS/cm to 1999 mS/cm) and process temperature range (-10 to 130oC, -10 to 105oC for ISC40SJ-TT) without changing the cell constant and conducting recalibration.

The erosion/abrasion resistant PEEK (Poly Ether Ether Ketone), which also features excellent chemical resistance in all solutions except Fluoric Acid or Oxidizing Concen- trated Acids.

The PEEK sensor is provided with a rugged Stainless Steel mounting thread/nut/gasket combination for ultimate flexibility in installation using bulk head installation technique. There is also a wide range of holders and options available for reliable in-line or off-line installation with double O-ring seals for long service life of the sensor.

The ISC40GJ, ISC40SJ have a large bore for optimal resistance to fouling processes and when properly installed, the flow will keep the sensor clean, to help avoid measuring errors.

 

FEATURES

Yokogawa General Purpose Inductive Conductivity Sensor ISC40GJ-GG-T1-10/SSG

• Inductive Conductivity technique for elimination of fouling and polarization errors.

• Wide bore sensors for long term stability.

• Installation flexibility due to wide range of holders and due to the use of universal bulkhead construction.

• A single sensor can maintain the high resolution and accuracy, and measure the conductivity in an extremely brood range.

  Minimum span: 100 μS/cm

  Maximum span: 1999 mS/cm

 

APPLICATIONS

Yokogawa General Purpose Inductive Conductivity Sensor ISC40GJ-GG-T1-10/SSG

• All applications where severe electrode fouling prevents the use of contacting electrodes.

• All ranges except (ultra) pure water applications.

• All slurry applications where conventional systems suffer from plugging or erosion.

• All applications where the 6 decade rangeability is necessary for accurate process control.

 

GENERAL SPECIFICATION 

Yokogawa General Purpose Inductive Conductivity Sensor ISC40GJ-GG-T1-10/SSG

1. ISC40GJ, ISC40SJ Inductive Conductivity Sensor

   Compatibility:

   ISC40GJ is compatible with FLXA21 2-wire Analyzer, ISC202G Inductive Conductivity Transmitter and ISC450G Converter.

   ISC40SJ-TT is compatible with ISC202SJ Inductive Conductivity Transmitter.

   Measuring range: 1 μS/cm to 1999 ms/cm

   Output span: Minimum 100 μS/cm

                           Maximum 1999 mS/cm

  Process temperature:

      -10 to 130 C for continuous exposure.

      -10 to 105 C for ISC40SJ-TT. Suitable for steam-sterilisable applications.

 Process pressure:

   Dependent on installation; but <2 MPa (300 psi).

   (Note) Process temperature and pressure depend on specification of holders and adapters.

 Process flow: Maximum 5 m/s.

 Wetted materials:

   Sensor: PEEK (Poly Ether Ether Ketone).

   O-ring: Viton® or ethylene propylene copolymer rubber.

   Adapter (optional): SUS316 Stainless steel or PVC or PVDF.

 Non-wetted materials:

    Sensor thread: SUS304 Stainless steel.

    Retaining nut:  SUS304 Stainless steel.

    Cable: Weatherproof vinyl.

 Process connection:

    With retaining nut on G3/4 thread of sensor top (refer to section Drawings and Dimensions)      for bulkhead mounting; optional process adapters      or process fittings.

  Process adapters:

    JIS 10K 50 RF flange adapter (Material: SUS316)

    JIS 10K 50 FF flange adapter (Material: PVC or PVDF)

    DIN PN16 DN50 flange adapter

    ANSI Class 150 2 flange adapter

    R2 screw-in adapter

 Cable length: 5 m, 10 m, 15 m, 20 m

  The length into extension cable is inside of 50 m

  Extension cable cannot be used with ISC40SJ-TT.

 Dimensions: Refer to section Drawings and Dimensions.

 Weight:   Sensor: approximately 0.6 kg.

 (Note)  Do not submerge the sensor itself in process water, as the seams between the mold and the metal of the sensor are not waterproof. Since a temperature sensor is imbedded in the PEEK molded sensor, its response speed is not fast. Install another temperature sensor if necessary.

 ISC40SJ-TT Intrinsically safe type sensor

 TIIS certification sensor should be used with ISC202SJ Protection Concept and Adapter Group: Ex ia IIC T4 Intrinsic safe rating:

Ui=14.4 V, Ii=20 mA, Pi=190 mW, Li=28.6 mH, Ci=0.2 μF

 Environment and operational conditions

 The temperature of the sample solution in contact with sensor should be the range of -10 to 105 C.

 

2. ISC40FDJ Immersion Holder

  Process temperature: Maximum 80 C.

  Process pressure: Maximum 0.2 MPa at 20 C.

                                   Maximum 0.1 MPa at 80 C. Wetted materials:

  Holder: C-PVC or SUS316

  O-ring: Viton® or ethylene propylene copolymer rubber. 

  Flange (Optional) : PP or SUS316

  Gasket: Chloroprene or ethylene propylene copolymer rubber.

  Process connection:

   Fixing flange (Optional): DIN PN10 DN50 (ANSI 2 inch 150 lbs. with bolt holes): Material PP

  JIS 10K 50 RF: Material SUS316 2-inch pipe Mounting set (Optional): Zinc-plated steel.

 

3. ISC40FFJ Flow Holder

  Process temperature:

  ISC40FFJ-SA, ISC40FFJ-SJ:  Maximum 150 C.

  ISC40FFJ-PA, ISC40FFJ-PJ:  Maximum 100 C.

  ISC40FFJ-FA, ISC40FFJ-FJ:  Maximum 130 C.

 Process pressure:

  ISC40FFJ-SA, ISC40FFJ-SJ: Maximum 1.0 MPa at 150 C.

  ISC40FFJ-PA, ISC40FFJ-PJ: Maximum 0.6 MPa at 20 C.

       ISC40FFJ-FA, ISC40FFJ-FJ: Maximum 0.1 MPa at 100 C.

                                                           Maximum 1.0 MPa at 20 C.

                                                           Maximum 0.1 MPa at 130 C.

 Wetted materials:

    ISC40FFJ-S口:    SUS316

    ISC40FFJ-P口:    Polypropylene

    ISC40FFJ-F口:    PVDF

  O-ring:  Viton® or ethylene propylene copolymer rubber.

 Non-wetted materials:

  Nut: SUS304

  Mounting set (Optional): SUS304

  Flange adapters (Optional): SUS304

Process connection:

1/2NPT or Rc1/2

DIN PN10 DN25 flange adapters (Optional)

JIS 10K 25 RF flange adapters (Optional)


4. ISC40FSJ Direct Insertion Subassembly

Process temperature:

  ISC40FSJ-STWJ : Maximum 110 C.

  ISC40FSJ-SCWJ, -SCSJ: Maximum 150 C.

  ISC40FSJ-PCSJ : Maximum 100 C.

Process pressure:

ISC40FSJ-SCWJ, -SCSJ: Maximum 1.0 MPa at 150 C.

ISC40FSJ-PCSJ: Maximum 0.6 MPa at 20 C.

                                 Maximum 0.1 MPa at 100 C.

ISC40FSJ-FCSJ: Maximum 1.0 MPa at 20 C.

Maximum 0.1 MPa at 130 C.

Materials:

Wetted materials:

ISC40FSJ-STWJ: SUS316L, silicon rubber.

ISC40FSJ-SCWJ, -SCSJ: SUS316, Viton® or ethylene propylene copolymer rubber.

ISC40FSJ-PCSJ: Polypropylene, Viton® or ethylene propylene copolymer rubber.

ISC40FSJ-FCSJ: PVDF, Viton® or ethylene propylene copolymer rubber.

Non wetted materials:

       ISC40FSJ-STWJ : IDF clamp: SUS13. 

       ISC40FSJ-SCWJ, -SCSJ, -PCSJ, -FCSJ: Nut: SUS304.

Process connection:

  ISC40FSJ-SCWJ: coupling.

  ISC40FSJ-SCSJ-PCSJ-FCSJ: R2 screw-in coupling.

Dimensions:

  Refer to section Drawings and Dimensions.

 

5. BA20 Terminal Box

Use when ISC202/FLXA21 transmitter or ISC450G converter is separated from ISC40口J sensor and is set up.

Ambient temperature: -10 To 50 C

Constructions: IP54 agreement

Case material: Article of cast metal of aluminum alloy

Cable inlet: 2 (Pg13.5) 

Case color : Straight gray 

Weight : Approx. 2 kg

(Note) BA20 can not be used with ISC40SJ-TT.

 

6. WF10J Extension Cable

Number of mind Line: 6

Finish outside diameter: 7.7 mm

Terminal processing: Special terminals

Material: Weatherproof vinyl.

    (Note) WF10J cannot be used with ISC40SJ-TT.

    Yokogawa General Purpose Inductive Conductivity Sensor ISC40GJ-GG-T1-10/SSG