Yokogawa Stop Valve L9852CB for Calibration gas line
General specifications
6. ZO21R Probe Protector
Used when sample gas flow velocity is approx. 10m/ sec or more and dust particles wears the detector in cases such as pulverized coal boiler of fluidized bed furnace (or burner) to protect the detector from wearing by dust particles.
Insertion Length: 1.05 m
Flange: JIS 5K 65A FF equivalent. ANSI Class
150 4 FF (without serration) equivalent or DIN PN10 DN50 A equivalent. However, flange thickness is different.
Material: SUS316 (JIS), SUS304 (JIS) (Flange)
Weight: 1.05 m; Approx. 6/10/8.5 kg (JIS/ANSI/DlN)
Installation: Bolts, nuts, and washers are provided for detector, probe adapter and process-side flange.
7. ZO21S Standard Gas Unit (*)
Standard Gas Unit must not be located in hazardous area.
Function: Portable unit for calibration gas supply consisting of span gas (air) pump, zero gas cylinder with sealed inlet, flow rate checker and flow rate needle valve.
Sealed Zero Gas Cylinders (6 provided): E7050BA
Capacity: 1 l
Filled pressure: Approx. 686 kPa G (at 35 °C)
Composition: 0.95 to 1.0 vol% O2+N2-balance
Power Supply: 100, 110, 115, 200, 220, 240V AC±10%, 50/60 Hz
Power Consumption: Max.5 VA
Case Material: SPCC (Cold rolled steel sheet)
Paint Color: Mainframe; Munsell 2.0 GY3.1/0.5 equivalent
Cover; Munsell 2.8 GY6.4/0.9 equivalent
Piping: Ø 6/Ø 4 mm flexible tube connection
Weight: Approx. 3 kg
(*) Non CE Mark.
8. ZA8F Flow Setting Unit
Used when instrument air is provided.
This unit consists of flowmeters and flow control valves to control the flow rates of calibration gas and reference gas.
FIowmeter Scale: Calibration gas; 0.1 to 1.0 l/min.
Reference gas; 0.1 to 1.0 l/min.
Construction: Dust-proof and rainproof construction
Case Material: SPCC (Cold rolled steel sheet)
Painting: Baked epoxy resin, Dark-green (Munsell 2.0 GY 3.1/0.5 or equivalent)
Tube Connections: Rc1/4 or 1/4 NPT (Female)
Reference Gas Pressure: Clean air supply of sample gas pressure plus approx. 50 kPa G (or sample gas pressure plus approx.150kPa when a check valve is used.).
Pressure at inlet of the Flow Setting Unit. (Max. 300 kPa G).
Air Consumption: Approx. 1.5 l/min
Weight: Approx. 2.3 kg
9. ZR40H Automatic Calibration Unit for Separate type Oxygen Analyzer
Yokogawa Stop Valve L9852CB for Calibration gas line
Automatic Calibration Unit must be located in Non- hazardous area.
Used when automatic calibration is required for the separate type and instrument air is provided. The solenoid valves are provided as standard.
Construction: Dust-proof and rainproof construction; NEMA 4X/IP67 - only for case coating solenoid valve, not flowmeter (excluding flowmeter)
Mounting: 2-inch pipe or wall mounting, no vibration
Materials: Body; Aluminum alloy, Piping; SUS316 (JIS), SUS304 (JIS), Flowmeter; MA (Methacrylate resin). Bracket; SUS304 (JIS)
Finish: Polyurethane corrosion-resistance coating, Mint green (Munsell 5.6BG3.3/2.9)
iping Connection: Rc1/4 or 1/4 NPT (Female)
Power Supply: 24 V DC (from ZR402G)
Power consumption: Approx. 1.3 W
Reference Gas Pressure: Sample gas pressure plus Approx. 150 kPa.
Pressure at inlet of automatic calibration unit. (690 kPa max.)
Air Consumption: Approx. 1.5 l/min
Weight: Approx. 3.5 kg
Ambient Temperature: -20 to +55°C, no condensing or freezing
Ambient Humidity: 0 to 95%RH Storage Temperature: -30 to +65°C
10. Automatic Calibration Unit for Integrated type Oxygen Analyzer
Yokogawa Stop Valve L9852CB for Calibration gas line
When Automatic Calibration of (-A) or (-B) code is specified, Automatic Calibration Unit is installed in the ZR202S.
Only Automatic Calibration Unit is not available.
11. L9852CB/G7016XH Stop Valve
The stop valve and the nipple are mounted on the calibration gas line.
The nipple is used to connect the stop valve. They are attached when the option code (/SV) is selected for the ZR22S or the ZR202S.
Connection: Rc1/4 (L9852CB) or 1/4 NPT (F) (G7016XH)
Material: SUS316 (JIS) Weight: Approx. 80 g
12. K9292DN/K9292DS Check Valve
This is used to prevent entry of sample gas into calibration gas line. Purpose is the same as stop valve, but is convenient, as it does not need to be opened or closed for calibration.
Mount directly on calibration gas inlet of detector in place of stop valve. However as source pressure of 150 kPa G or more is needed, standard gas unit cannot be used. When option code “/CV” of the ZR22S or the ZR202S is specified, check valve is provided.
Connection: Rc1/4 (K9292DN) or 1/4 NPT (F) (K9292DS)
Material: SUS304 (JIS)
Pressure: 70 kPa G or more and 350 kPa G or less
Weight: Approx. 40g
13. Air Set
Yokogawa Stop Valve L9852CB for Calibration gas line
This set is used to lower the pressure when instrument air is used as the reference and span gases.
G7003XF/K9473XK
Primary Pressure: Max. 1 MPa G
Secondary Pressure: 0.02 to 0.2 MPa G
Connection: Rc1/4 or 1/4 NPT (F) with joint adapter
Weight: Approx. 1 kg
G7004XF/K9473XG
Primary Pressure: Max. 1 MPa G
Secondary Pressure: 0.02 to 0.5 MPa G
Connection: Rc1/4 or 1/4 NPT (F) with joint adapter
Weight: Approx. 1 kg
14. G7013XF/G7014XF Pressure Regulator
Primary Pressure: 14.8 MPa G, Secondary Pressure: 0 to 0.4 MPa G
Connection: Inlet; W22 14 threads, right hand screw
Outlet; Rc1/4 or 1/4 NPT (Female) Material: Brass body
15. ZR22A, ZR202A Heater Assembly
ZR22A: Spare Parts for ZR22S
ZR202A: Spare Parts for ZR202S
(Note) Yokogawa shall not guarantee the heater assembly after its replacement.
16. E7044KF Case Assembly of Calibration Gas Cylinder
Installation: 2B pipe mounting
Material: SPCC (Cold rolled steel sheet)
Case Paint: Baked epoxy resin, Jade green (Munsell 7.5 BG 4/1.5)
Weight: Approx. 10 kg with gas cylinder
(Note) Export of such high pressure filled gas cylinders to most countries is prohibited or restricted.
Model and Suffix Codes
Yokogawa Stop Valve L9852CB for Calibration gas line
Automatic Calibration Unit for Separate type Oxygen Analyzer
Model
|
Suffix code
|
Option code
|
Description
|
ZR40H
|
- - - - - - -
|
- - - -
|
Automatic calibration
unit for ZR402G
|
Gas piping connection
|
-R
-T
|
- - - - - -
|
Rc1/4
1/4 NPT (F)
|
Wiring
connection
|
-P
-G
-M
-T
|
- - - - - -
- - - - - -
|
Pipe connection (G1/2)
Pg 13.5
M20 x 1.5
1/2 NPT
|
─
|
-A
|
- - - -
|
Always -A
|