General Specifications:
Cable with pin terminals (applicable to FLXA21,SC202G and SC202SJ)
Cable with M3 ring terminals (applicable to SC450G,)
Cable with fork terminals (applicable to SC202G and SC202口/TB)
Cable with M4 ring terminals (applicable to FLXA21)
Object of measurement: Conductivity of liquids
Measuring Principle: 2-electrode system or 4-electrode system
Cell Constants: 0.01 cm-1 or 10 cm-1 (for two-electrode system)
10 cm-1 (for four-electrode system)
Temperature Range: 0˚ to 100˚C (130˚C maximum only for 0.01 cm-1 cell constant detectors, excluding those with polypropylene chambers)
Pressure: 1000 kPa max. (500 kPa maxi- mum for detectors with polypropy- lene chambers)
Flow rate of Sample Solution: No particular limitation applies, although a value of less than 20 l/ min. is recommended for flow- through detectors.
(Note) No limitation appliesto flow rate (flow velocity) as far as measurement is concerned. Take care, however, when usingflow-through detectors. Electrodes or the inner wallsof a liquid chamber may wear put drastically at higher flow speeds if a measured solution contains slurry. Air bubblesshould not be mixedin the sample solutions to obtain correct measured values.
RTD for Temperature Compensation:
Yokogawa Conductivity Sensor SC8SG
Pt1000 (built into the sensor) Construction : Rainproof encapsulation (compatible with the JIS C0920 Japanese Industrial Standard)
Weight:
Yokogawa Conductivity Sensor SC8SG
Screw-in type: approximately 1.3 kg (excluding the cable)
Flow-through type (SCS14 chamber): approximately 3.1 kg (excluding the cable)
Flow-through type (SCS14 chamber, flanged): approximately 4.5 kg (excluding the cable)
Flow-through type (polypropylene chamber): approximately 2.7 kg (excluding the cable)
Flow-through type (polypropylene chamber, flanged): approximately 3.2 kg (excluding the cable)
Cable: 0.3 kg for 5.5-m length; 0.5 kg for 10-m length; 0.9 kg for 20-m length.
Process Connection: Screw-in or flow-through
Construction of Wetted Part:
Yokogawa Conductivity Sensor SC8SG
• Sensor-holding base:
SUS316 and fluoro-rubber when using screw- in type holder or the chamber made of stainless steel. PP and fluoro-rubber when using the chamber made of PP.
• 0.01 cm-1 cell constant, two-electrode sensor:
SUS316 and ethylene chloride trifluoride
• 10 cm-1 cell constant, two-electrode sensor:
reinforced epoxy resin and graphite
• 10 cm-1 cell constant, four-electrode sensor:
polyvinylidene difluoride, glass and platinum
• Stem (flow-through type):
SCS14 or polypropylene resin
Installation:
Yokogawa Conductivity Sensor SC8SG
• Screw-in type—held by the process piping
• Flow-through type (polypropylene chamber)
—mounted on a pipe (nominal diameter of 50 mm 2 in.)
• Flow-through type (SCS14 chamber)
—held by the process piping
Spare Parts for SC8SG
Yokogawa Conductivity Sensor SC8SG
Parts
No.
|
Description
|
K9208BA
K9208BC
K9208BD
K9208BK G9303EB
|
0.01cm-1 cell constant, two-electrode sensor
10cm-1 cell constant, two-electrode sensor
10cm-1 cell constant, four-electrode
sensor
Welding socket for screw-in model
O-ring
|