Yokogawa Indicator YS1310-140/A04
GENERAL
Yokogawa Indicator YS1310-140/A04
The YS1310-140/A04 Indicator with Alarms is an input indicator that has alarms with two inputs. Two loops can be displayed simultaneously. A high limit alarm, low limit alarm, high-high limit alarm, and low-low limit alarm can be generated for each of the two inputs, and the AND connection or OR connection of arbitrary alarms can be output from a total of 6 contacts.
The YS1310-140/A04 offers high reliability thanks to Yokogawa’s proprietary technology, user friendliness, and expandabil- ity. The basic type conforming to the DIN short case size contributes to space saving for mounting, and complies with standards (CE, FM* and CSA*). For easy replace- ment of earlier models requiring the same panel cutout dimensions and depth as those of earlier models are also offered. *: To be approved.
FEATURES
Yokogawa Indicator YS1310-140/A04
▪ A full-dot, FTF LCD Clear visibility of the display screen is ensured even in direct sunlight in the early morning and late afternoon.
The user can freely access a desired operation screen from meter, trend, bar graph, alarm, and event displays.
All parameters can be set via the front panel display.
▪ The parameters can all be set by operation on the front panel (no need to draw out the internal unit).
▪ Parameters can also be set using YSS1000 Setting Software available separately.
▪ Can be driven by either an AC (100 V) or DC (24 V) power supply. (Must be specified upon ordering if using a 220–240 V DC power supply.)
▪ Dust- and splash-proof IP54 faceplate (for basic type only)
▪ Depth 250 mm (only for the basic type)
▪ No battery or capacitor is used for memory backup. No maintenance is required because batteries and backup capacitors are not used.
▪ CE mark (for basic types and YS100-compatible type only)
▪ FM non-incentive explosion protection (option, for basic types only) (Under an application) (To be approved)
▪ CSA non-incentive explosion protection (option, for basic type, compatible type for YS100 (with YS100 case)) (To be approved)
▪ Communication (option)
Ethernet (Modbus/TCP) (for basic types only)
RS485 (PC Link, Modbus, YS protocol, unavailable for YS80 internal unit-compatible type)
▪ Compatibility with the YS100 series
Setting and control operations can be done with the same feel. In the case of the basic type, terminal-to-terminal pitches differ but the signal-to-terminal arrangement is almost the same. The following functions are included to ensure compatibility with the SIHK.
▪ Meter Display was added to the Operation Display.
▪ Housing for replacement purposes (SHUP-000 sepa- rately available)
Model and Suffix Code
Yokogawa Indicator YS1310-140/A04
Model
|
Suffix Code
|
Optional Code
|
Remark
|
YS1310-140/A04
|
|
Indicator with alarm
|
Use
|
-1
|
Always -1
|
Type
|
0
|
Basic type
|
2
|
YS100 compatible type (with YS100 case)
|
3
|
YS80 internal unit compatible type, EBS, I, EK, HOMAC compatible type (*1)
|
4
|
YS80 compatible type (YS 80 compatible size, YS100 terminal type)
|
5
|
100 line compatible type (YS100 terminal type) (*2)
|
Power supply
|
0
|
100 VAC and 24 VDC power supply
|
1
|
220 VAC power supply
|
Direct input (*3)
|
/A01
|
mV input
|
/A02
|
Thermocouple input
|
/A03
|
Resistance temperature detector input
|
/A04
|
Potentiometer input
|
/A05
|
Isolator
|
/A06
|
2-wire transmitter input (insulation type)
|
/A07
|
2-wire transmitter input (non-insulation type)
|
/A08
|
Frequency input
|
/DF
|
Direct input with Fahrenheit temperature range function (*7)
|
Communication
|
/A31
|
RS-485 communication (PC link, Modbus, YS protocol) (*4)
|
/A34
|
Ethernet communication (Modbus/TCP) (*5)
|
Certification
|
/FM
|
FM nonincendive approved (FM Class I, Division 2) (*6) (To be approved)
|
/CSA
|
CSA safety and nonincendive approved (Class I, Division 2) (*8) (To be approved)
|
|
|
|
|
|
|
*1: Can be connected with the YS80 housing (Model: SHUP-000). (The EK and HOMAC compatible housing SHUP-420 and EBS/I series compatible housing SHUP-100 are available separately.)
*2: The 100 line compatible housing (Model: YS006) is available separately.
*3: The direct input option can be added only for suffix codes –12x, -14x, and -15x. Multiple selections are not possible.
*4: Can not be added for suffix code –13x.
*5: /A34 can be specified only for suffix code –10x.
*6: Can be added only for suffix code –10x.
*7 This option can be combined only with option code /A02 or /A03.If option code /DF is specified, Fahrenheit temperature range can be available for direct
input range in addition to Centigrade temperature range. In case of specifying Fahrenheit temperature range for direct input, option code /DF is required.
When the direct input temperature range may be changed to Fahrenheit temperature range after shipment, also specify option code /DF.
*8 Can be added only for suffix code –10x and –12x.
Display and Setting Functions
Yokogawa Indicator YS1310-140/A04
● Display functions
Display specification
YS1700 displays are composed of the following three groups and the individual functions can be set up via displays for the respective settings:
Operation displays
|
LOOP displays TREND displays ALARM displays DUAL displays METER displays
FAIL display (when a failure occurs)
|
Tuning displays
|
Setting display
Input/output data display
|
Engineering
|
Function settings
Input specification settings Password setting Operation display settings LCD settings Communication settings
|