Yokogawa Paperless Recorder FX1010-4-2-L
GENERAL SPECIFICATIONS
Yokogawa Paperless Recorder FX1010-4-2-L
Construction
Mounting:
|
Flush panel mounting (on a vertical plane) Mounting may be inclined downward up to 30 degrees from a horizontal plane.
|
Allowable panel thickness:
|
2 to 26 mm
|
Front panel:
|
Water- and dustproof: Complies with IEC529-IP65 (excluding side-by-side mounting)
|
Input
Yokogawa Paperless Recorder FX1010-4-2-L
Number of inputs:
|
FX1002: 2 channels, FX1004: 4 channels, FX1006: 6 channels, FX1008: 8 channels, FX1010: 10 channels, FX1012: 12 channels
|
Measurement intervals:
|
FX1002, DX1004: 125 ms, 250 ms FX1006, FX1008, FX1010, FX1012: 1 s, 2 s, 5 s
|
Inputs:
|
DCV (20, 60, 200 mV, 1, 2, 6, 20, 50 V, 1-5 V) TC (R, S, B, K, E, J, T, N, W, L, U, WRe) RTD (Pt100, JPt100) DI (Contact input, TTL level) DCA (with external shunt resistor attached)
|
Measurement/display accuracy:
|
Standard operating conditions: Temperature: 23 ± 2°C Humidity: 55% ± 10%RH Power supply voltage: 90 to 132 or 180 to 250 VAC Power supply frequency: 50/60 Hz ± 1% Warm-up time: At least 30 minutes. Other ambient conditions such as vibration should not adversely affect the operation.
|
Input
|
Range
|
Measurement Accuracy
|
Digital Display Max. Resolution
|
DCV
|
1-5 V
|
±(0.05% of rdg+3 digits)
|
1 mV
|
Thermocouple*
|
K
|
±(0.15% of rdg + 0.7°C) –200 to –100°C: ±(0.15% of rdg + 1°C)
|
0.1°C
|
RTD
|
Pt100
|
±(0.15% of rdg+0.3°C) 0.1°C
|
0.1°C
|
* Does not include the accuracy of reference junction compensation
Display
Yokogawa Paperless Recorder FX1010-4-2-L
Display:
|
5.7-inch TFT color LCD (240 × 320 dots) * A section of the LCD monitor may contain pixels that are always on or off. The brightness of the LCD may also not be uniform due to the characteristics of the LCD. This is not a malfunction.
|
Display groups:
|
Number of groups: 10 Number of channels that can be assigned to each group: Up to 6
|
Display color:
|
Channel: Select from 24 colors Background: White or black (selectable)
|
Trend display:
|
Layout: Vertical, horizontal, or wide
|
Bar graph display:
|
Direction: Vertical or horizontal (selectable)
|
Digital display:
|
Update rate: 1 sec
|
Overview display:
|
Measuring values and alarm status of all channels
|
Information display:
|
Alarm summary, message summary, memory summary, report, stacked bar graph, status, Modbus status
|
Modbus log display:
|
Displays the login log, error log, communication log (/C2, /C3, and /C7), FTP log (/C7), Web log (/C7), e-mail log (/C7), SNTP log (/C7), and DHCP log (/C7)
|
Tag display:
|
Number of displayable characters: Up to 16 Displayable characters: English, Japanese, and Chinese
|
Messages:
|
Number of displayable characters: Up to 32 alphanumeric Displayable characters: English, Japanese, and Chinese
|
Historical display function:
|
Plays back data from internal memory or external memory media.
|
Back light saver function:
|
LCD back light dims or turns OFF (user selectable) if no keys have been pressed for a specified period (1, 2, 5, 10, 30, or 60 min).
|
Data Saving Function
Yokogawa Paperless Recorder FX1010-4-2-L
External storage medium:
|
Medium: CompactFlash memory card (CF card) (on FXs with a CF card slot)
|
Internal memory:
|
Medium: Flash memory Format: FAT32 or FAT16 Capacity: 400 MB Maximum number of files that can be saved: 400 (total number of display data and event data files) Operation: FIFO (First In, First Out)
|
Alarm Function
Number of alarm levels:
|
Up to four for each channel
|
Alarm types:
|
High limit, low limit, differential high limit, difference low limit, high rate-of-change limit, low rate-of-change limit, alarm delay high limit, and alarm delay low limit
|
Event Action Function
General:
|
A particular action can be executed by particular event.
|
Number of event actions:
|
40 actions can be set
|
Security Functions
Yokogawa Paperless Recorder FX1010-4-2-L
General:
|
Login function or key lock function can be set for each key operation or communication operation.
|
Key lock function:
|
On/off and password can be set for each operation key and FUNC operation.
|
Login function:
|
User name and password to login can be set. System administrators: 5 (with access to all operations) Users: 30 (with access to operations based on their user access rights)
|
Clock
Clock:
|
With calendar function (Western calendar)
|
Accuracy:
|
±50 ppm (0 to 50°C); does not include the delay (1 second or less) that occurs when the power is turned on.
|
Batch Function
General:
|
Data display and data management with batch name, text field function and batch comment function are available.
|
Power Supply
Yokogawa Paperless Recorder FX1010-4-2-L
Rated power supply:
|
100 to 240 VAC (automatic switching)
|
Allowable power supply voltage range:
|
90 to 132 or 180 to 264 VAC
|
Rated power supply frequency:
|
50/60 Hz (automatic switching)
|
Power consumption:
|
Max. 45 VA (for 240 VAC power supply)
|
NORMAL OPERATING CONDITIONS
Supply voltage:
|
90 to 132, 180 to 250 VAC
|
Rated power supply frequency:
|
50 Hz ±2%, 60 Hz ±2%
|
Ambient temperature:
|
0 to 50°C
|
Ambient humidity:
|
20 to 80% RH (at 5 to 40°C), 10 to 50% (at 40 to 50°C)
|
Model and Suffix Code
Yokogawa Paperless Recorder FX1010-4-2-L
Model code
|
Suffix code
|
Optional code
|
Description
|
FX1002
|
|
|
|
|
2ch, Shortest measurement interval: 125ms
|
FX1004
|
|
|
|
|
4ch, Shortest measurement interval: 125ms
|
FX1006
|
|
|
|
|
6ch, Shortest measurement interval: 1s
|
FX1008
|
|
|
|
|
8ch, Shortest measurement interval: 1s
|
FX1010
|
|
|
|
|
10ch, Shortest measurement interval: 1s
|
FX1012
|
|
|
|
|
12ch, Shortest measurement interval: 1s
|
External storage medium slot
|
-0
|
|
|
|
Without CF card slot and medium (Note)
|
-4
|
|
|
|
With CF card slot and medium
|
Language
|
-2
|
|
|
English/Japanese/German/French/Chinese/Italian/Spanish/Portuguese/ Russian/Korean, deg F and DST
|
Withstanding voltage between measuring input terminals
|
-H
|
|
1000 VAC(50/60 Hz), 1 min
|
-L
|
|
400 VAC(50/60 Hz), 1 min
|
Options
|
/A1
|
Alarm output 2 points (C-contact)*1*10
|
/A2
|
Alarm output 4 points (C-contact)*1
|
/A3
|
Alarm output 6 points (C-contact)*1*3
|
/A4A
|
Alarm output 12 points (A-contact)*1*3
|
/C2
|
RS-232 interface*2
|
/C3
|
RS-422A/485 interface*2
|
/C7
|
Ethernet interface
|
/F1
|
FAIL/Status output*3
|
/M1
|
Mathematical functions (including Report functions)
|
/N2
|
3 leg isolated RTD*4
|
/N3F
|
Extended input type (without Pt1000)
|
/P1
|
24 VDC/AC power supply
|
/R1
|
Remote control 8 points*5
|
/TPS2
|
24VDC transmitter power supply (2 loops)*6*10
|
/TPS4
|
24VDC transmitter power suply (4 loops)*7
|
/USB1
|
USB interface (1 port)
|
/PM1
|
Pulse input 3 points, Remote control 5 points (including Mathematical functions)*8
|
/CC1
|
Calibration correction function
|
/LG1
|
Log scale
|
/PWR1
|
Power monitor (including Mathmatical functions)*9*10
|
Note: To load data, the FX must be equipped
with a communication interface (/C2, /C3, or /C7 option) or the USB interface (/USB1 option).
*1 Any combination of /A1, /A2, /A3, and /A4A cannot be specified together.
*2 /C2 and /C3 cannot be specified together.
*3 /A3 or /A4A cannot be specified together with /F1.
*4 /N2 cannot be specified for FX1002 or FX1004.
*5 If /R1 is specified, /A4A, /TPS2, /TPS4, /PM1, or /PWR1 cannot be specified.
*6 If /TPS2 is specified, /TPS4, /A2, /A3, /A4A, /F1, /R1, or /PM1 cannot be specified.
*7 If /TPS4 is specified, /TPS2, /A1, /A2, /A3, /A4A, /F1, /R1, or /PM1 cannot be specified.
*8 If /PM1 is specified, /A4A, /M1, /R1, /TPS2, /TPS4, or /PWR1 cannot be specified.
*9 If /PWR1 is specified, /A3, /A4A, /F1, /R1, /PM1, or /M1 cannot be specified.
*10 /TPS2, /PWR1, and /A1 cannot be specified together.